Study of the sensitivity of culture media after bio decontamination for use in an isolator

14 October 2019

B. Palas1, H. Doisneau2, H. Balkhi1, H. Doillet1, C. Jaskowiec1, N. Fauchet2, S. Poullain1
1 Pharmacy Department, Centre Hospitalier Intercommunal de Créteil, France
2 Department of Bacteriology and Hygiene, Centre Hospitalier Intercommunal de Créteil, France

Context
Our production of parenteral nutrition and chemotherapy mixtures has been carried out under isolator since the summer of 2018. In accordance with good preparation practices, we carry out microbiological controls of the environment regularly. In addition, the sterility of each batch of parenteral nutrition is controlled by seeding at least 4 pairs of blood cultures.

Objective
We want to determine the viability of culture media after bio decontamination with hydrogen peroxide, if there is a significant difference in microbiological growth after bio decontamination of surface agar plates, air agar plates and blood culture vials.

Material and method
For blood cultures we seeded 8 pairs (aerobic + anaerobic) of vials (type BactAlert®), with 10 CFU of S. aureus then 50 CFU, on control vials and vials subjected to hydrogen peroxide during 22min (> 100ppm), totaling 32 pairs. The vials were put in the incubator (Virtuo®) and the "positivity" delay was recorded. For air and surface agar plates, 8 pairs of agar plates of each type (control and peroxide, BioMérieux® brand, TSA plate, simple and without protection against the sterilizing agent) were inoculated with approximately 25 CFU of the following germs: E. coli, S. aureus, P.aeruginosa, C. albicans, M. luteus, 80 air and 80 contact. After 24 to 48 hours in a microbiological incubator at optimal temperatures for each medium, we counted the number of CFU of each agar plate.

Results
For blood cultures, the study did not show a significant difference (threshold of 5%) between control vials and vials bio-decontaminated at 2 concentrations of germs for the "positivity" delay (p<0.001), 12.6h to 10 CFU and 10.6h to 50 CFU. For air agar, there is a decrease in sensitivity after biodecontamination for all germs (p <0.001) except C. albicans. For surface agar plates, there is a decrease in sensitivity after bio-decontamination for all germs (p <0.001) except for C. albicans and M. luteus.
Decreases in sensitivity are dependent on agar and germs.

Conclusion
air and surface agar plates are no longer usable after exposure to hydrogen peroxide. There is no impact for BactAlert® blood cultures vials. There are several alternatives marketed: triple-packaged agar, screw agar or agar with sterilizing agent inhibitor. These results show the importance of checking the conditions of use of the culture media.

See the poster

Discussion forum

GERPAC provides for its members a discussion forum specifically devoted to technology in hospital pharmacy

Access forum

GERPAC
Association Loi 1901
Siège social : Chez Jean-Yves Jomier / 8ter rue Léon Bussat, 64000 PAU
Immatriculation formation N° 72 64 035 30 64
Contact us Legal notes - Management of personal data
Réalisation studiomaiis.net : Développeur web WordPress & SPIP freelance